About this trial
Effective ventilation during cardiopulmonary resuscitation is essential but can be difficult to achieve consistently. Current practice relies on visible chest rise to guide ventilation volume, although this may not always accurately reflect the actual volume of air delivered. This study aims to determine whether a real-time ventilation feedback device, providing objective information on ventilation volume and air leakage, improves adherence to recommended ventilation volumes compared with standard ventilation without feedback. The study will focus on patients receiving CPR with a supraglottic airway and synchronous ventilation.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
None
Disqualifiers
None
Trial design
Parallel
Treatments tested in this trial
Ventilation Feedback Device (Active)
DeviceThe same flow sensor and device measure inspiratory and expiratory tidal volume, leakage, and ventilation timing, connected between the resuscitator bag and airway device. In feedback mode, the device displays real-time visual feedback to guide the provider toward a target expiratory tidal volume of 400-500 mL.
Control (Recording-Only)
DeviceA ventilation feedback device integrated into standard resuscitation equipment, used during out-of-hospital cardiac arrest. The device measures inspiratory and expiratory tidal volume, leakage, and ventilation timing via a flow sensor connected between the resuscitator bag and airway device. In this arm, the device is randomly initiated in recording-only mode, recording the same ventilation parameters without displaying any feedback to the provider.
Treatment groups
Trial outcomes
Primary outcomes
Proportion of guideline-adherent ventilation pauses per resuscitation attempt
A ventilation pause is a compression pause during CPR in which rescue breaths are delivered. Guideline adherence per pause is defined as delivery of two ventilations, each with a device-measured expiratory tidal volume of 400-500 mL, per European Resuscitation Council recommendations. For each patient, the proportion of guideline-adherent pauses is calculated as the number of adherent pauses divided by the total analyzable pauses during the resuscitation attempt. The primary outcome is the mean of this patient-level proportion, averaged across all patients, compared between feedback and control groups.
Secondary outcomes
Number of Participants with Prehospital Return of Spontaneous Circulation (ROSC)
Any return of spontaneous circulation (ROSC), defined per the Utstein Out-of-Hospital Cardiac Arrest Registry Template, at any point during the prehospital resuscitation attempt.
Number of Participants Surviving to 30 Days
Survival to 30 days
Number of Participants with Favorable Neurological Outcome
Favorable neurological outcome, defined as a Modified Rankin Scale (mRS) score of 0-3, where 0 indicates no symptoms and 6 indicates death, assessed at hospital discharge or 30 days.
Proportion of Guideline-Adherent Ventilations
Proportion (%) of ventilations with expiratory tidal volume 400-500 mL, per European Resuscitation Council recommendations, calculated across all ventilations pooled across patients during prehospital resuscitation. Device-recorded data, compared between feedback and control groups.
Other outcomes
Change in Ventilation Quality Over Time Within Resuscitation Attempts
Trend in the proportion (percentage-point change) of guideline-adherent ventilations across consecutive time intervals within a resuscitation attempt, pooled across patients. Device-recorded, compared between feedback and control groups.
EMS-Level Learning Curve in Ventilation Quality
Change in ventilation quality, expressed as percentage-point change in guideline adherence, at the EMS-provider level as a function of cumulative study exposure (number of cases performed). Device-recorded.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Jeroen van Eijk
Lead sponsor
Amsterdam UMC, location VUmc
Sponsor institution
Archeon Medical