About this trial
This pilot feasibility study will evaluate UltraTrac, an investigational noninvasive wearable device for continuous cardiac output monitoring, in patients with congenital heart disease.
UltraTrac integrates impedance cardiography and carotid Doppler ultrasound to estimate cardiac output. Participants undergoing clinically indicated cardiac catheterization, cardiac MRI, or cardiac intensive care unit monitoring will have UltraTrac measurements collected during routine clinical care.
UltraTrac measurements will be compared with reference standard or clinical hemodynamic measurements, including thermodilution, direct Fick method, cardiac magnetic resonance imaging-derived cardiac output, and clinical hemodynamic assessments before and after vasoactive medication changes. UltraTrac data will not be used to guide clinical decision-making.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
0-80 years old
Participant weight of at least 2 kg
Diagnosis of congenital heart disease of any anatomic type, including but not limited to repaired or unrepaired single ventricle physiology, biventricular congenital heart disease, valvular disease, septal defects, and conotruncal anomalies
Clinically indicated cardiac catheterization, cardiac MRI, or cardiac intensive care unit admission during the study enrollment period
Disqualifiers
Known allergy or prior skin reaction to adhesive electrodes or ultrasound gel
Skin breakdown, wound, or infection at any proposed electrode or probe placement site that would preclude safe device application
Hemodynamic instability at the time of enrollment that, in the judgment of the treating clinician, would preclude safe participation in research procedures
Known carotid artery abnormality, prior carotid surgery, or anatomic variant that would preclude reliable carotid Doppler signal acquisition
Trial design
Parallel
Treatments tested in this trial
UltraTrac
DeviceUltraTrac is an investigational wearable hemodynamic monitoring device that integrates impedance cardiography and carotid Doppler ultrasound into a single noninvasive platform. In this study, UltraTrac will be used only as an observational monitoring device. UltraTrac measurements will not guide clinical decision-making, and all clinical management will proceed according to standard of care.
Treatment groups
Trial outcomes
Primary outcomes
Agreement Between UltraTrac-Derived Cardiac Output and Reference Standard Cardiac Output During Cardiac Catheterization
Agreement between cardiac output measurements obtained periprocedural by UltraTrac and cardiac output measurements obtained by thermodilution and/or direct Fick method during clinically indicated cardiac catheterization will be assessed using Bland-Altman analysis. Percentage error will be calculated as two standard deviations of the bias divided by the mean cardiac output of the two methods. A percentage error of 30% or less will be considered clinically acceptable agreement.
Secondary outcomes
Agreement Between UltraTrac-Derived Cardiac Output and CMR-Derived Cardiac Output
Periprocedural agreement between UltraTrac-derived and CMR-derived cardiac output (L/min) and cardiac index (L/min/m²) will be assessed using Bland-Altman analysis. CMR-derived cardiac output will be obtained during clinically indicated cardiac MRI using phase contrast flow measurement and volumetric assessment. Each participant will contribute one measurement pair per encounter, consisting of one CMR-derived cardiac output value and the temporally closest UltraTrac-derived measurement. Reported parameters will include mean bias, limits of agreement, percentage error, and 95% confidence intervals for the bias and limits of agreement. Percentage error will be calculated as 2 standard deviations of the bias divided by the mean cardiac output of the two methods × 100. Percentage error ≤30% will indicate clinically acceptable agreement by Critchley and Critchley criteria. Results will be reported overall and by age group, weight category, and cardiac anatomy.
Concordance Rate Between UltraTrac-Derived Directional Cardiac Output Changes and Clinical Hemodynamic Assessment in the CICU
The ability of UltraTrac to detect directional changes in cardiac output will be assessed during the immediate postoperative period and throughout the broader CICU admission. Measurements will be obtained during pre-defined windows before and after each clinically indicated vasoactive medication initiation or titration event. The composite clinical reference standard will include heart rate, arterial blood pressure, capillary refill time, acid-base status and lactate, and invasive hemodynamic parameters where available (arterial line, central venous pressure). Concordance between UltraTrac-derived directional cardiac output changes and composite clinical hemodynamic assessment will be reported as a concordance rate (%). The specific statistical method for concordance analysis will be determined and pre-specified prior to data collection and will be detailed in the statistical analysis plan.
Percentage of Participants With Successful UltraTrac Device Placement and Analyzable Waveform Acquisition
Feasibility of UltraTrac device placement and signal acquisition will be assessed as the proportion of enrolled participants with successful device placement and analyzable waveform data acquisition across the cardiac catheterization laboratory, cardiac MRI suite, and cardiac intensive care unit settings.
Time Required for UltraTrac Device Application
Time required for UltraTrac device application will be recorded by study personnel during the specific clinical setting in which the device is applied.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
This trial is not recruiting at the moment. You can still explore other options: