Video Laryngeal Mask SaCoVLM Versus LMA Supreme and Ambu AuraGain in Children

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age1-10
SponsorMasarykova Nemocnice v Usti nad Labem, Krajska Zdravotni a.s.

About this trial

This randomised controlled trial compares three supraglottic airway devices (laryngeal masks) in children with a body weight of 3-30 kg undergoing elective surgery under general anaesthesia: the SaCoVLM video laryngeal mask (UE Medical), which has an integrated camera allowing visual confirmation and adjustment of mask position, the LMA Supreme (Teleflex) and the Ambu AuraGain (Ambu). Children are randomly assigned in a 1:1:1 ratio to one of the three devices. The main question is whether the SaCoVLM provides a higher oropharyngeal leak pressure (a measure of how well the mask seals the airway) than the LMA Supreme and the Ambu AuraGain. Secondary questions include first-attempt insertion success, the need for repositioning or reinsertion of the mask during surgery and airway-related complications.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Paediatric patients scheduled for elective surgery under general anaesthesia with a supraglottic airway device

Body weight 3-30 kg (supraglottic airway device sizes 1-2.5)

Written informed consent of the parent or legal guardian

Disqualifiers

Body weight < 3 kg or > 30 kg

Anticipated difficult airway

History of difficult airway

Indication for airway management by a method other than a supraglottic airway device

Trial design

Design model

Parallel

Treatments tested in this trial

  • SaCoVLM video laryngeal mask

    Device

    Airway management with the SaCoVLM video laryngeal mask (UE Medical, China), a second-generation supraglottic airway device with an integrated camera. The video display is used to visualise the laryngeal inlet and to verify and, if necessary, adjust the device position. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

  • LMA Supreme

    Device

    Airway management with the LMA Supreme (Teleflex, USA), a second-generation supraglottic airway device inserted using the standard blind technique. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

  • Ambu AuraGain

    Device

    Airway management with the Ambu AuraGain (Ambu A/S, Denmark), a second-generation supraglottic airway device inserted using the standard blind technique. Size per manufacturer weight-based recommendation; cuff pressure 40-60 cmH2O.

Treatment groups

135 Participants
are divided into 3 treatment groups
Group A: SaCoVLMExperimental treatment 1 intervention
Group B: LMA SupremeActive comparator 1 intervention
Group C: Ambu AuraGainActive comparator 1 intervention

Trial outcomes

Primary outcomes

1

Oropharyngeal leak pressure (OLP)

Airway pressure (cmH2O) at which the circuit manometer reading becomes stable, measured with the breathing circuit in manual mode, ventilation paused, fresh gas flow 10 L/min and the APL valve set to 50 cmH2O, after stable ventilation is achieved.

Time frame
From device insertion up to 5 minutes after insertion

Secondary outcomes

1

First-attempt insertion success rate

Proportion of patients in whom the device is successfully inserted with adequate ventilation at the first attempt.

Time frame
From the start of the first insertion attempt until successful insertion (maximum 3 attempts), up to 10 minutes
2

Need for device repositioning or reinsertion during surgery

Proportion of patients requiring repositioning, reinsertion of the device or other airway intervention after successful insertion.

Time frame
From device insertion until device removal at the end of anaesthesia, up to 5 hours
3

Incidence of perioperative airway-related complications

Proportion of patients with at least one of: desaturation (SpO2 \< 92%), laryngospasm, bronchospasm, aspiration; each complication also reported separately.

Time frame
From device insertion until device removal at the end of anaesthesia, up to 5 hours
4

Blood staining of the device on removal

Proportion of patients with visible blood on the device or in secretions after device removal.

Time frame
At device removal at the end of emergence from anaesthesia, up to 5 hours after insertion

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Masarykova Nemocnice v Usti nad Labem, Krajska Zdravotni a.s.

Lead sponsor

Faculty of Health Studies, J. E. Purkyně University in Ústí nad Labem

Collaborator

Charles University, Czech Republic