About this trial
This study evaluates EMILY.AI, an investigational (non-CE-marked) smart oxygen valve that automatically adjusts the oxygen flow delivered to a patient.
Adults with chronic respiratory disease who need oxygen during exertion perform two six-minute walk tests on the same day, 30 minutes apart: one with oxygen adjusted automatically by EMILY.AI and one with a fixed oxygen flow delivered through the same device. The order is randomized and participants do not know which test uses automatic titration.
The study compares the time spent within the target saturation range, desaturation, oxygen use, walking distance, breathlessness and participant preference between the two tests. This substudy is part of clinical investigation FJD-EMILY-23-02, whose other substudies are registered in separate records.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Aged 40 to 90 years with a chronic respiratory disease.
Indication for exertional oxygen therapy, defined as SpO2 below 88% during a six-minute walk test and/or resting respiratory failure (PaO2 below 60 mmHg).
Able to understand the study and sign informed consent.
Disqualifiers
Decompensated acute or chronic disease with severe organ involvement, such as cardiovascular disease.
Musculoskeletal or neurological disease preventing stable and safe walking.
Comorbidities or treatments that, by intensity or dose, limit performance of the tests or interpretation of the results (such as alcohol use, beta-blockers, or other neurological or psychiatric drugs).
Walking limitation preventing the six-minute walk test, or cardiovascular or connective tissue disease limiting accurate transcutaneous pulse oximetry.
Trial design
Crossover
Treatments tested in this trial
EMILY.AI automatic oxygen titration
DeviceInvestigational smart oxygen valve with rule-based algorithm that adjusts oxygen flow every second from pulse-oximetry SpO2, heart rate and respiratory rate during walking and recovery.
Fixed-flow oxygen
Other interventionStandard care: constant oxygen flow set by the physician, delivered through the EMILY.AI device in fixed mode so that the participant cannot distinguish it from automatic titration.
Treatment groups
Trial outcomes
Primary outcomes
Percentage of time with SpO2 within the target range
Percentage of test time (6 minutes of walking plus 2 minutes of recovery) with SpO2, measured by pulse oximetry, within the predefined target range (88-92%), without delivering oxygen while SpO2 is above the range. Compared between the test with automatic titration by EMILY.AI and the test with fixed flow.
Secondary outcomes
Percentage of test time with SpO2 below 88%
Percentage of test time (walking and recovery) with SpO2 below 88%, compared between conditions.
Number of desaturation episodes with SpO2 below 88%
Number of episodes per test in which SpO2 falls below 88%, compared between conditions.
Time to start of oxygen flow correction
Time in seconds from SpO2 leaving the target range (88-96%) to the start of an oxygen flow change, compared between conditions.
Time to return of SpO2 to the target range
Time in seconds from SpO2 leaving the target range (88-96%) to its return to the range, compared between conditions.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
Aether Tech S.L.
Lead sponsor
Hospital Universitario Fundación Jiménez Díaz
Collaborator