About this trial
Cataract is the leading cause of reversible blindness and the first surgery worldwide. It corresponds to the crystalline lens opacification. Nevertheless, its grading remains mostly subjective and the Lens Opacities Classification System dating back 1993, is not used in daily practice and has not been revisited since. The HYPERLAF device developed by the start-up HYPEREYE is a modified slit lamp which combines imaging and spectral analysis, providing two quantitative objective scores: OPAK (opacity) and COLOR (spectral analysis). It is a non-CE marked prototype with non-invasive imaging methods.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Patients enrolled in or eligible for a social security program
Age 18 or older
Patients with cataracts. In cases of bilateral cataracts, both eyes will be photographed and analyzed.
Patient who has already received an instillation of 0.5% mydriaticum (tropicamide) and 10% neosynephrine (Théa Laboratories, Clermont-Ferrand, France) eye drops to dilate the pupil of the eye to be photographed (or both eyes).
Disqualifiers
Severe blepharospasm making the examination impossible.
Poor or absent pupil dilation: patient with an allergy or intolerance to the eye drops used for pupil dilation; refusal of pupil dilation during the eye examination.
Pregnant or breastfeeding patient
person under legal guardianship
Trial design
Single group
Treatments tested in this trial
Non-contact observation of the crystalline lens nucleus
Diagnostic testNon-contact observation of the crystalline lens nucleus with a modified slit lamp combining an imaging system and a spectral measurement.
Treatment groups
Trial outcomes
Primary outcomes
Reproducibility of COLOR score measurement.
To assess the reproducibility of the COLOR score measurement, we will perform three consecutive measurements for each lens in each subject. In practice, the measurement procedure will be repeated from the beginning: for each measurement, the patient will sit on the LAF, the observer will focus the light on the lens, take the measurement, and then move on to the other eye. At the end, the patient will step back. Reproducibility will be quantified using the intraclass correlation coefficient (ICC).
Secondary outcomes
Reproducibility of OPAK score
To assess the reproducibility of the OPAK score measurement, we will perform three consecutive measurements for each lens in each subject. In practice, the measurement procedure will be repeated from the beginning: for each measurement, the patient will sit on the LAF, the observer will focus the light on the lens, take the measurement, and then move on to the other eye. At the end, the patient will step back. Reproducibility will be quantified using the intraclass correlation coefficient (ICC).
Distribution of the COLOR score in the study population
The COLOR score values measured in the study population will be described using appropriate descriptive statistics, including distribution, variability, and minimum and maximum values.
Distribution of the OPAK score in the study population
The OPAK score values measured in the study population will be described using appropriate descriptive statistics, including distribution, variability, and minimum and maximum values.
Association between the COLOR score and LOCS III components
The association between the COLOR score and the individual components of the LOCS III (Lens Opacities Classification System III) score will be analyzed using appropriate statistical methods. LOCS III components will be assessed from digital lens photographs obtained using the LAF as part of routine care in the Ophthalmology Department of Saint-Etienne University Hospital.
Sponsors and contacts
Click on the lead sponsor to view all of their trials.
This trial is not recruiting at the moment. You can still explore other options: