Performance Evaluation of the OCUTOP Hand-held Corneal Topographer

Trial statusNot yet recruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexAll
Age18+
SponsorUniversity of Plymouth

About this trial

The OCUTOP is a new hand-held camera that photographs the front of the eye to map the shape of the cornea and to assess the tear film. Before it can be used in eye care it must be compared with the instruments that clinics already use. In this study, adults attending one of two UK eye clinics for an ordinary appointment are invited to have both eyes measured once with the OCUTOP and with the clinic's usual instruments, in a random order, and to complete two short questionnaires. The visit adds about 30 minutes; nothing is put in the eye and no treatment is given. The study measures how closely the OCUTOP agrees with the reference instruments for corneal curvature, tear break-up time and tear meniscus height, how repeatable it is, how often it produces usable images, how acceptable participants and clinicians find it, and whether any adverse events occur.

Eligibility criteria

This trial accepts healthy volunteers

Qualifiers

Aged 18 years or over.

Able to give written informed consent.

Attending the participating clinic for any reason.

Able to sit and fixate on a target for the duration of the measurements.

Disqualifiers

Unable or unwilling to give informed consent.

Active infection or inflammation of the eye or eyelids, or broken skin around the eye, at the study visit.

Any condition that prevents the eye being kept open or the head being kept steady for the measurements (for example severe blepharospasm or nystagmus).

Contact lens wear on the day of the visit.

Trial design

Design model

Single group

Treatments tested in this trial

  • OCUTOP hand-held corneal topographer with OCUGRAPH software

    Device

    Investigational hand-held, non-contact, Placido-based corneal topographer with integrated non-invasive tear-film assessment (tear break-up time and tear meniscus height), operated through the OCUGRAPH software on a Windows computer. The device rim rests on the skin around the eye; nothing touches the eye. Both eyes are imaged at one visit, with repeat captures for repeatability. The device is not UKCA or CE marked and is labelled exclusively for clinical investigation. Manufacturer: OCUWELL Limited.

  • Reference instrument measurement (corneal topography and tear-film assessment)

    Diagnostic test

    Corneal topography and non-invasive tear-film assessment with the CE or UKCA marked instruments in routine use at each site, used within their intended purpose: Pentacam HR (topography) and Keratograph 5M (tear film) at Site 01, with the Medmont E300 as an additional topographer; Myah (topography and tear film) at Site 02. Both eyes, same visit, repeat captures for repeatability.

Treatment groups

150 Participants
are divided into 1 treatment group
Group A: OCUTOP and reference instrumentsExperimental treatment 2 interventions

Trial outcomes

Primary outcomes

1

Agreement between the OCUTOP and the reference instrument for central simulated mean keratometry (K-mean)

Mean paired difference (OCUTOP minus reference, dioptres) with 95% confidence interval, and Bland-Altman 95% limits of agreement with their 95% confidence intervals, in the analysis eye, pooled across sites and by site. Reference: Pentacam HR (Site 01), Myah (Site 02).

Time frame
Single study visit (Day 1)
2

Agreement between the OCUTOP and the reference instrument for first non-invasive tear break-up time (NIBUT-first)

Mean paired difference (OCUTOP minus reference, seconds) with 95% confidence interval, and Bland-Altman 95% limits of agreement with their 95% confidence intervals, in the analysis eye, pooled across sites and by site. Reference: Keratograph 5M (Site 01), Myah (Site 02).

Time frame
Single study visit (Day 1)
3

Agreement between the OCUTOP and the reference instrument for average non-invasive tear break-up time (NIBUT-average)

Mean paired difference (OCUTOP minus reference, seconds) with 95% confidence interval, and Bland-Altman 95% limits of agreement with their 95% confidence intervals, in the analysis eye, pooled across sites and by site. Reference: Keratograph 5M (Site 01), Myah (Site 02).

Time frame
Single study visit (Day 1)
4

Agreement between the OCUTOP and the reference instrument for central lower tear meniscus height (TMH)

Mean paired difference (OCUTOP minus reference, millimetres) with 95% confidence interval, and Bland-Altman 95% limits of agreement with their 95% confidence intervals, in the analysis eye, pooled across sites and by site. Reference: Keratograph 5M (Site 01), Myah (Site 02).

Time frame
Single study visit (Day 1)

Secondary outcomes

1

Repeatability of the OCUTOP and of each reference instrument

Within-subject standard deviation, repeatability coefficient, coefficient of variation and intraclass correlation coefficient from repeat captures, for K-mean, NIBUT-first, NIBUT-average, TMH, K1, K2 and the astigmatism vectors J0 and J45.

Time frame
Single study visit (Day 1)
2

Agreement for K1, K2 and the astigmatism vectors J0 and J45

Mean paired difference and 95% limits of agreement with 95% confidence intervals for K1, K2, J0 and J45, OCUTOP against each reference instrument and, at Site 01, against the Medmont E300.

Time frame
Single study visit (Day 1)
3

Classification agreement at TFOS cut-offs

Agreement between the OCUTOP and the reference instrument in classifying eyes at the TFOS cut-offs (NIBUT below 10 s; TMH below 0.2 mm): Cohen's kappa, sensitivity and specificity with the reference as standard.

Time frame
Single study visit (Day 1)
4

Inter-operator reproducibility of the OCUTOP

Between-operator differences and limits of agreement for the primary measures in the first 20 participants per site, each measured with the OCUTOP by two operators.

Time frame
Single study visit (Day 1)

Other outcomes