Portable Absence Seizure Detection Using Smart Earbuds

Trial statusNot yet recruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age6+
SponsorUniversitaire Ziekenhuizen KU Leuven

About this trial

Typical absence seizures are the hallmark of idiopathic generalized epilepsy syndromes (CAE, JAE, JME) but are frequently underreported in patient diaries. Clinical studies have shown that 74% to 94% of absence seizures go unreported by patients or caregivers. While 24-hour video-EEG (vEEG) is the gold standard for quantification, it is expensive, hospital-based, and lacks the temporal resolution required for long-term management.

Previous studies successfully validated a wearable EEG (wEEG) solution using a behind-the-ear (BTE) form factor (Byteflies Sensor Dot). In the pilot study, the authors demonstrated that a reduced electrode configuration combined with ML algorithms could detect absence seizures with 98% sensitivity in the hospital setting and enable effective home-based treatment monitoring.

To expand the applicability of this method, validation of additional form factors is required. This study aims to extend our validation framework to a novel wEEG device that utilizes an ear configuration (earEEG).

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Voluntary written informed consent of the participant or their legally authorized representative has been obtained prior to any screening procedures.

At least 18 years of age at the time of signing the ICF or at least 6 years of age at the time their parents/legal representatives sign the caregiver's IFC and the minor participants sign the age-appropriate IFC.

Confirmed diagnosis of the following idiopathic generalized epilepsy syndromes: CAE, JAE, or JME.

Use of highly effective methods of birth control; defined as those that, alone or in combination, result in low failure rate (i.e., less than 1% per year) when used consistently and correctly; such as implants, injectables, combined oral contraceptives, some Intra-Uterine Devices (IUDs), true sexual abstinence (i.e. refraining from heterosexual intercourse during the entire period of risk associated with the investigation treatment(s)) or commitment to a vasectomised partner.

Disqualifiers

Patient has a history of other generalized epilepsy syndromes, such as Lennox-Gastaut.

Anatomical Constraints: Anatomical abnormalities of the outer ear or ear canal that prevent the proper fit of the earEEG device.

Participation in another clinical trial with an investigational drug or device that would interfere with the endpoints of this study.

Female who is pregnant, breastfeeding, or intends to become pregnant or is of child-bearing potential and not using an adequate, highly effective contraceptive

Trial population

Adults and children with idiopathic generalized epilepsy syndromes and typical absence seizures.

Trial design

Design model

Case-only

Time perspective

Prospective

Treatments tested in this trial

Not listed

Trial groups

25 Participants
are grouped into 1 trial group
Group A: Absence patients

Trial outcomes

Primary outcomes

1

Sensitivity (Recall)

The proportion of true absence seizures (defined as 3Hz Spike and Wave Discharges ≥ 3 seconds identified on the gold-standard vEEG) that are correctly detected by the wEEG algorithm.

Time frame
1 day, during simultaneous use of the IMD and 25-channel scalp EEG (in-hospital phase)
2

Precision

The proportion of positive predictions that were actually correct when compared to the vEEG.

Time frame
1 day, during simultaneous use of the IMD and 25-channel scalp EEG (in-hospital phase)
3

F1-score

The harmonic mean metric used to evaluate the overall performance of the ML model. It uses both precision and sensitivity.

Time frame
1 day, during simultaneous use of the IMD and 25-channel scalp EEG (in-hospital phase)
4

False Alarm Rate in 24 hours

The number of false positive detections per 24 hours of recording.

Time frame
1 day, during simultaneous use of the IMD and 25-channel scalp EEG (in-hospital phase)

Secondary outcomes

1

Signal quality

Comparative hourly analysis of root mean square (RMS) and signal-to-noise ratio in the alpha band (SNR-Alpha) in the signals originated with the IMD channels and the spatially corresponding channels of the standard 10-20 EEG.

Time frame
1 day, during simultaneous use of the IMD and 25-channel scalp EEG (in-hospital phase)
2

Usability Scores

Quantitative assessment of user experience (patients/caregivers) using standardized questionnaires (i.e., System Usability Scale, final questionnaire-comfort \& user experience, and User Perspective Questionnaire).

Time frame
3 weeks, when the IMD will be used during 24 hours continuously at least once a week.
3

Data Yield

The percentage of interpretable (artifact-free) data obtained during unsupervised home recording sessions.

Time frame
3 weeks, when the IMD will be used during 24 hours continuously at least once a week.