About this trial
To correlate ocular morphology from anterior segment OCT and pyramid wavefront sensing in order to comprehensively analyze pseudophakic in vivo image quality of presbyopia-correcting IOLs in cataract patients.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Men and women aged above 21 years with prior uncomplicated cataract/IOL exchange surgery
Willing to sign informed consent before measurements
Corrected distance visual acuity (CDVA) of 0.1 logMAR or better
Absence of corneal scars/dry eye disease that could falsify the light transmission of the study eye
Disqualifiers
Prior corneal refractive or ocular surgery
Posterior capsule opacification.
Keratometry between <42.5 and >45 D,
Corneal spherical aberration > 0.4 microns 5) Angle kappa > 0.4 mm
Trial population
80 cataract patients (160 eyes)
Trial design
Cohort
Prospective
Treatments tested in this trial
Osiris-T Pyramid Wavefront Sensor - (CSO, Italy)
DeviceThe Osiris-T aberrometer is a CE-certified device that operates via a high-resolution four-faced pyramid wavefront sensor (PWS) to analyze light intensity among four images (sub-apertures) of the patients ́entrance pupil akin to a Foucault knife-edge test. Thereby measurements of the wavefront and refractive error can be quantified non-invasively. The sampling density of the wavefront error is 45,000 points at maximum pupil. The device uses a near-infrared extended light-emitting diode source emitting at 850 nm. Its is additionally intertwined with a corneal topography system to also characterize the corneal aberrations and compute via direct subtraction the internal ocular optics.
Treatment groups
9
Treatment groupsSee each treatment group below.
Trial outcomes
Primary outcomes
IOL ocular optics measument
To measure the internal ocular optics of patients with presbyopia-correcting IOLs via 3-D anterior segment OCT biometry and pyramid wavefront sensing by employing the following optical metrics higher-order aberrations (HOAs, unit: root means square)
Secondary outcomes
Comparison of clinical optical quality and device measurements
Correlate and compare clinical optical quality via defocus curves (by employing trial lenses to the best distance prescription, starting from +0.50 to -3.00 diopters (D), in -0.50-D steps) to virtual ray traced aberrations from anterior segment OCT as well as physically measured wavefront aberrations with the pyramid wavefront sensor. Wavefront data will thereby be also employed to analyze the fidelity of virtual ray traced aberrations from anterior segment OCT and measured wavefront aberrations with the pyramid wavefront sensor.
Higher-order aberrations in non-dilated and dilated pupils
Changes in higher-order aberrations (unit: root mean square) will be evaluated before and after pharmacological dilation of pupils
Non-dilated and dilated Pupil centration
Changes in pupil centration (unit: mm) will be evaluated before and after pharmacological dilation
Correlation and Comparison of IOL decentration
Correlate and compare IOL decentration measured with the pyramid-wavefront sensor to IOL decentration and tilt measured with 3-D anterior segment OCT.