RIGHTEST iFree 2 CGM in Pregnant Women With Diabetes Study

Trial statusRecruiting
Trial phaseNot applicable
Trial typeInterventional
Biological sexFemale
Age18+
SponsorBionime Corporation

About this trial

This study will evaluate the accuracy and safety of the RIGHTEST iFree 2 Continuous Glucose Monitoring (CGM) System in pregnant women with type 1 diabetes, type 2 diabetes, or gestational diabetes. The system uses a small sensor placed under the skin to measure glucose levels in the fluid between cells.

Approximately 104 women will participate at study sites in Germany and Taiwan. Each participant will wear two sensors, one on each upper arm, for up to 15 days. During one clinic visit lasting up to six hours, blood samples will be collected every 15 minutes. Sensor readings will be compared with laboratory blood glucose measurements to assess accuracy.

Participants will also be assessed for side effects, including skin reactions at the sensor sites, and will receive a safety follow-up after sensor removal. No procedures will be used to deliberately raise or lower blood glucose levels.

Eligibility criteria

This trial does not accept healthy volunteers

Qualifiers

Age 18 years or older at enrollment.

Diagnosis of type 1 diabetes mellitus, type 2 diabetes mellitus, or gestational diabetes mellitus.

Gestational age of at least 12 weeks.

Singleton pregnancy.

Disqualifiers

Known allergy to, or inability to tolerate, medical-grade adhesive tape.

Extensive unresolved skin changes or diseases at sensor wear sites that prevent sensor placement, including extensive psoriasis, recent burns or severe sunburn, extensive eczema, scarring, tattoos, or dermatitis herpetiformis.

Hematocrit below 30%.

Hemoglobin below 11.0 g/dL in the first or third trimester, or below 10.5 g/dL in the second trimester, at screening.

Trial design

Design model

Single group

Treatments tested in this trial

  • RIGHTEST iFree 2 Continuous Glucose Monitoring System

    Device

    The RIGHTEST iFree 2 Continuous Glucose Monitoring System (Bionime Corporation) measures interstitial glucose continuously for up to 15 days.

Treatment groups

104 Participants
are divided into 1 treatment group
Group A: RIGHTEST iFree 2 CGM SystemExperimental treatment 1 intervention

Trial outcomes

Primary outcomes

1

Percentage of CGM Readings Within ±20% of Reference Blood Glucose Measurements

Percentage of raw CGM readings from the designated primary sensor within ±20% of paired venous blood glucose measurements obtained using the YSI 2300 analyzer, across the device measuring range. CGM and reference measurements are paired within a maximum time difference of 5 minutes.

Time frame
During one sampling session of up to 6 hours on Day 1, Day 5 ±1, Day 10 ±1, or Day 15 of sensor wear.

Secondary outcomes

1

CGM Accuracy

Accuracy of primary-sensor CGM readings against paired YSI 2300 blood glucose measurements, including agreement within ±20% after user calibration; 15/15%, 20/20%, and 40/40% agreement overall and by glucose range; glucose-range concurrence; mean absolute relative difference (MARD) by glucose range; reference values corresponding to LOW/HIGH readings; Clarke and Consensus Error Grid zone distributions; and glucose rate-of-change concurrence. Glucose ranges include \<54, 54-69, 70-180, 181-250, and \>250 mg/dL. Raw and user-calibrated readings are analyzed separately.

Time frame
During one sampling session of up to 6 hours on Day 1, Day 5 ±1, Day 10 ±1, or Day 15 of sensor wear.
2

Between-Sensor Precision

Precision assessed using mean paired absolute relative difference (PARD) and coefficient of variation (CV) between two sensors worn simultaneously by each participant. Readings are paired within ±30 seconds. Raw and user-calibrated readings are analyzed separately.

Time frame
Throughout the sensor wear period, up to 15 days.
3

Performance Across the Sensor Wear Period

Primary-sensor accuracy assessed by 15/15%, 20/20%, and 40/40% agreement and MARD at Day 1, Day 5 ±1, Day 10 ±1, and Day 15, with each participant contributing one assigned sampling visit. Raw and user-calibrated readings are analyzed separately. Sensor survival and reading provision rates are assessed throughout wear using both primary and secondary sensors.

Time frame
Throughout sensor wear, up to 15 days; reference sampling on one assigned day: Day 1, Day 5 ±1, Day 10 ±1, or Day 15.
4

Hypoglycemia and Hyperglycemia Alert Performance

True alert, false alert, correct detection, and missed detection rates for primary-sensor hypoglycemia and hyperglycemia alerts. Alerts are compared with reference blood glucose measurements using a ±15-minute matching window around device alerts or reference glucose events.

Time frame
During one sampling session of up to 6 hours on Day 1, Day 5 ±1, Day 10 ±1, or Day 15 of sensor wear.

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.