About this trial
This study will evaluate the accuracy and safety of the RIGHTEST iFree 2 Continuous Glucose Monitoring (CGM) System in pregnant women with type 1 diabetes, type 2 diabetes, or gestational diabetes. The system uses a small sensor placed under the skin to measure glucose levels in the fluid between cells.
Approximately 104 women will participate at study sites in Germany and Taiwan. Each participant will wear two sensors, one on each upper arm, for up to 15 days. During one clinic visit lasting up to six hours, blood samples will be collected every 15 minutes. Sensor readings will be compared with laboratory blood glucose measurements to assess accuracy.
Participants will also be assessed for side effects, including skin reactions at the sensor sites, and will receive a safety follow-up after sensor removal. No procedures will be used to deliberately raise or lower blood glucose levels.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Age 18 years or older at enrollment.
Diagnosis of type 1 diabetes mellitus, type 2 diabetes mellitus, or gestational diabetes mellitus.
Gestational age of at least 12 weeks.
Singleton pregnancy.
Disqualifiers
Known allergy to, or inability to tolerate, medical-grade adhesive tape.
Extensive unresolved skin changes or diseases at sensor wear sites that prevent sensor placement, including extensive psoriasis, recent burns or severe sunburn, extensive eczema, scarring, tattoos, or dermatitis herpetiformis.
Hematocrit below 30%.
Hemoglobin below 11.0 g/dL in the first or third trimester, or below 10.5 g/dL in the second trimester, at screening.
Trial design
Single group
Treatments tested in this trial
RIGHTEST iFree 2 Continuous Glucose Monitoring System
DeviceThe RIGHTEST iFree 2 Continuous Glucose Monitoring System (Bionime Corporation) measures interstitial glucose continuously for up to 15 days.
Treatment groups
Trial outcomes
Primary outcomes
Percentage of CGM Readings Within ±20% of Reference Blood Glucose Measurements
Percentage of raw CGM readings from the designated primary sensor within ±20% of paired venous blood glucose measurements obtained using the YSI 2300 analyzer, across the device measuring range. CGM and reference measurements are paired within a maximum time difference of 5 minutes.
Secondary outcomes
CGM Accuracy
Accuracy of primary-sensor CGM readings against paired YSI 2300 blood glucose measurements, including agreement within ±20% after user calibration; 15/15%, 20/20%, and 40/40% agreement overall and by glucose range; glucose-range concurrence; mean absolute relative difference (MARD) by glucose range; reference values corresponding to LOW/HIGH readings; Clarke and Consensus Error Grid zone distributions; and glucose rate-of-change concurrence. Glucose ranges include \<54, 54-69, 70-180, 181-250, and \>250 mg/dL. Raw and user-calibrated readings are analyzed separately.
Between-Sensor Precision
Precision assessed using mean paired absolute relative difference (PARD) and coefficient of variation (CV) between two sensors worn simultaneously by each participant. Readings are paired within ±30 seconds. Raw and user-calibrated readings are analyzed separately.
Performance Across the Sensor Wear Period
Primary-sensor accuracy assessed by 15/15%, 20/20%, and 40/40% agreement and MARD at Day 1, Day 5 ±1, Day 10 ±1, and Day 15, with each participant contributing one assigned sampling visit. Raw and user-calibrated readings are analyzed separately. Sensor survival and reading provision rates are assessed throughout wear using both primary and secondary sensors.
Hypoglycemia and Hyperglycemia Alert Performance
True alert, false alert, correct detection, and missed detection rates for primary-sensor hypoglycemia and hyperglycemia alerts. Alerts are compared with reference blood glucose measurements using a ±15-minute matching window around device alerts or reference glucose events.